See Full Document Text
2022
Guidance on
Stability Studies of
In-vitro Diagnostic
Medical Device
(IVDMD)
Draft for public comment
Central Drugs Standard Control Organization
Directorate General of Health Services
Ministry of Health and Family Welfare
Government of IndiaCentral Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
Central Drugs Standard
Control Organization
(In-Vitro Diagnostic Division)
Guidance Document
Title : Guidance on Stability Studies of In-
Vitro Diagnostic Medical Device
(IVDMD)
Doc : CDSCO/IVD/GD/Stability/01/2022
No.
Date : 07.07.2022
Notice:
This Guidance document is aimed only for creating public awareness about In-Vitro Diagnostic
Devices Regulation by CDSCO and is not meant to be used for legal or professional purposes. The
readers are advised to refer to the statutory provisions of Medical Device Rules, 2017 and
subsequent amendments and clarifications issued by CDSCO time to time for all their professional
needs.
Page | 2Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
Table of Contents
Sr. No. Content Page No.
Preface 4
1. Introduction 5
2. Definitions and abbreviations 5
3. MDR 2017 requirements 7
4. Basic principles for stability testing 8
5. Shelf-life studies 12
6. Component stability studies 13
7. Stability during transport 16
8. In-use stability studies 17
9. Production lots used in stability studies 20
10. Stability plan 21
11. Stability report 25
12. Changes to a Licensed / Approved IVD 26
APPENDIX
APPENDIX I Example stability protocols. 29
APPENDIX II Specimens for the stability testing panels 36
APPENDIX III Examples of stability study approaches 41
Page | 3Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
Preface:
CDSCO is pleased to announce the release of the draft Guidance Document
on stability studies of In-vitro Diagnostic Medical Device (IVDMD). It is guidance for
Manufacturers in Preparation of a Premarket Review Document for Class C and
Class D IVDMD Import or Manufacturing License Applications.
This document is intended to aid manufacturers in the preparation of
scientific information to be provided in support of claimed shelf life, in use stability
and shipping studies for Class C and Class D IVDMD license applications and Post
approval change application filed in pursuant to the Medical Devices Rules, 2017
(MDR-2017). This document has been developed by the CDSCO to encourage and
support convergence of regulatory systems for medical Devices among
jurisdictions. CDSCO is looking to adopt the use of this Guidance for premarket
license applications and Post approval change applications. CDSCO strongly
encourages manufacturers to follow this guidance when Submitting Class C and
Class D IVDMD license applications and Post approval change applications.
This guidance document integrates global regulatory practices within the
Medical Devices Rules, 2017 (MDR-2017) licensing requirements for in vitro
diagnostic device license applications.
Please note that once implemented, all premarket in vitro diagnostic device
license applications are expected to be prepared as specified in this guidance.
The proposed guidance document is being uploaded for the information of all
stakeholders likely to be affected thereby for comments, if any.
Any person interested making any suggestions on the proposed draft
guidance documents may do so in writing for consideration of the CDSCO with in a
period of 30 days from the date of its uploading, through post to the Drugs
Controller General (India), CDSCO, FDA Bhavan, Kotla Road, New Delhi – 110002
and through email at ivd-division@cdsco.nic.in.
Page | 4Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
1. Introduction
1.1. Key concepts
Stability is the ability of an IVD reagent to maintain its performance
characteristics over a defined time interval. The purpose of stability studies
is to verify the time period and the storage conditions over which stable
performance characteristics of an IVD can be claimed.
1.2. Rationale of stability studies
The stability of an IVD is fundamental for its reliable performance over a
defined period of time. It is a regulatory requirement for the manufacturer to
provide objective, scientifically sound evidence to support all claims made
regarding the stability of an IVD. In addition a manufacturer can use stability
studies to show that all lots manufactured during the commercial life of the
IVD will meet predetermined user needs.
2. Definitions and abbreviations
2.1. Definitions
The definitions given below apply to the terms used in this document. They
may have different meaning in other contexts.
Accelerated stability evaluation:
Study designed to increase the rate of chemical and/or physical degradation,
or change, of an IVD reagent by using stress environmental conditions to
predict shelf-life.
NOTE: The design of an accelerated stability evaluation can include extreme
conditions of temperature, humidity, light or vibration.
Acceptance criteria:
A defined set of conditions that must be met to establish the performance of
a system. Numerical limits, ranges, or other suitable measures for
acceptance of the results of analytical procedures.
Accuracy of measurement:
Closeness of the agreement between the result of a measurement and a true
value of the measurand.
NOTE 1: Accuracy of measurement is related to both trueness of
measurement and precision of measurement.
NOTE 2: Accuracy cannot be given a numerical value in terms of the
measurand, only descriptions such as 'sufficient' or 'insufficient' for a stated
purpose.
Arrhenius plot:
Mathematical function that describes the approximate relationship between
the rate constant of a chemical reaction and the temperature and energy of
activation.
Batch/Lot:
Defined amount of material that is uniform in its properties and has been
produced in one process or series of processes.
Page | 5Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
Component:
Part of a finished, packaged and labelled IVD medical device.
NOTE : Typical kit components include antibody solutions, buffer solutions,
Calibrators and/or control materials
Constituent:
Raw materials used to make a component. The physical and performance
requirements of an IVD that are used as a basis for IVD design.
Drift:
Characteristic slow change of a metrological value from a measuring
instrument.
Environmental factors:
Variables that might affect the performance or efficacy of IVD reagents e.g.
temperature, airflow, humidity, light, this also includes dust and micro-
organisms.
Evidence:
Information which can be proved true, based on facts obtained through
observation, measurement, test or other means
Instructions for Use (IFU):
Information supplied by the manufacturer to enable the safe and proper use
of an IVD
NOTE: Includes the directions supplied by the manufacturer for the use,
maintenance, troubleshooting and disposal of an IVD, as well as warnings
and precautions.
In vitro diagnostic (IVD):
A medical device, whether used alone or in combination, intended by the
manufacturer for the in vitro examination of specimens derived from the
human body solely or principally to provide information for diagnostic,
monitoring or compatibility purposes.
NOTE: IVDs include reagents, calibrators, control materials, specimen
receptacles, used, for example, for the following test purposes: diagnosis, aid
to diagnosis, screening, monitoring, predisposition, prognosis, prediction,
determination of physiological status.
IVD reagent:
Chemical, biological or immunological components, solutions, or
preparations intended by the manufacturer to be used as an IVD.
Metrological traceability:
Property of the result of a measurement or the value of a standard whereby it
can be related to stated references, usually national or international
standards, through an unbroken chain of comparisons all having stated
uncertainties.
NOTE:- Each comparison is affected by a (reference) measurement
procedure defined in a calibration transfer protocol.
Page | 6Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
Performance claim:
Specification of a performance characteristic of an IVD as documented in the
information supplied by the manufacturer
NOTE:- This can be based upon prospective performance studies, available
performance data or studies published in the scientific literature.
“Information supplied by the manufacturer” includes but is not limited to:
statements in the IFU, in the dossier supplied to CDSCO and /or other
regulatory authorities.
3. MDR-2017 requirements:-
The domestic manufacturer or authorized agent, in case of Import, shall submit
the duly signed detailed information pertaining to stability of the product in Device
master file as specified in point 15-18 of Appendix III of fourth Schedule part II of
MDR 2017; Manufacturer should describe claimed shelf life, in use stability and
shipping studies, should provide information on stability testing studies to support
the claimed shelf life.
i. Claimed Shelf life:
This section should provide information on stability testing studies to support
the claimed shelf life. Testing should be performed on at least three different
lots manufactured under conditions that are essentially equivalent to routine
production conditions (these lots do not need to be consecutive lots).
Accelerated studies or extrapolated data from real time data are acceptable
for initial shelf life claim but need to be followed up with real time stability
studies. Such detailed information should describe:
(a) the study report (including the protocol, number of lots, acceptance
criteria and testing intervals);
(b) when accelerated studies have been performed in anticipation of the
real time studies, the method used for accelerated studies;
(c) conclusions and claimed shelf life.
Explanation- Shelf life can be derived from the lot with the longest real
time stability data as long as accelerated or extrapolated data from all
three lots are comparable.
ii. In use stability:
This section should provide information on in use stability studies for one lot
reflecting actual routine use of the device (real or simulated). This may include
open vial stability and/or, for automated instruments, on board stability. In the
case of automated instrumentation if calibration stability is claimed, supporting
data should be included. Such detailed information should describe:
(a) the study report (including the protocol, acceptance criteria and testing
intervals);
(b) conclusions and claimed in use stability.
iii. Shipping stability:
This section should provide information on shipping stability studies for one lot
to evaluate the tolerance of products to the anticipated shipping conditions.
Page | 7Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
Shipping studies can be done under real and/or simulated conditions and
should include variable shipping conditions such as extreme heat or cold.
Such information should describe:
(a) the study report (including the protocol, acceptance criteria);
(b) method used for simulated conditions;
(c) conclusion and recommended shipping conditions.
3.1. Manufacturer responsibility
It is a manufacturer‟s responsibility to ensure that all claims made regarding the
stability of the IVD performance are supported by objective, scientifically-sound
evidence.
3.2. Standards
CDSCO recommends the following standards for the use in establishment of
stability claims: ISO 23640:2013, CLSI EP25-A and ASTM:D4169-14. It is
recommended that manufacturers be familiar with these standards and consider
them when designing and planning their stability studies.
3.3. Suitability for use in India
The stability studies submitted to CDSCO should accurately reflect the
expected environmental conditions and the normal usage
conditions/methods encountered by the users in India‟s States, such as:
Extremes of temperature for in-use conditions and during
transportation
Extremes of humidity encountered during in-use conditions,
transportation and storage
Dust
Light, both the amount required for accurate testing/results
interpretation and any affects that light may have on the IVD
functionality
Micro-organisms
3.4. Meeting customer requirements
By undertaking well-designed stability studies including periodic verification
activities, the manufacturer can demonstrate that the product meets
customer requirements, as required by Fifth Schedule (Quality Management
System for medical devices and in vitro diagnostic medical devices) of MDR,
2017. Meeting predetermined user expectations, not merely evaluating the
capability of an IVD, is a fundamental aspect of development of IVDs. It is a
proactive means for the manufacturer to prevent quality problems at lot
release and in the post-production and marketing phase.
4. Basic principles for stability testing
4.1. Critical characteristics of the IVD
A well-designed stability study must generate evidence of stability of each of
the critical constituents of the IVD (risk-evaluated critical constituents), each
of the claimed analytes, and any particular level of performance including
precision, sensitivity and specificity of the kit.
Page | 8Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
Examples:
1) A hepatitis C virus (HCV) assay containing the critical constituents
related to detection of NS3 or core proteins should have the stability of all
such constituents proven.
2) For an assay designed to detect both IgG and IgM by use of protein A
and protein L, the stability of both protein A and protein L should be
proven.
3) For CD4, all the antibodies involved (e.g. anti-CD3 and anti-CD4) must
be shown to be stable.
4) For an IVD claimed to detect particular seroconversion specimens, or
genotypes, or to have specified precision at particular analyte
concentrations, or a particular specificity, each of these claims must be
proven over the stated shelf-life.
4.2. Finalized product presentation
During stability testing, all IVD components (including the device, calibrator
and / or control material, etc.) must be made and tested to the finalized
manufacturing documentation and in the finalized packaging including
intended labels and containers. All presentations (e.g. different buffer
volumes used for different kit sizes) must be used during stability testing.
4.3. Environmental conditions
The study should subject the IVD to a combination of conditions which
define the limits of stability for all lots made during its commercial life. The
combinations of conditions, durations of exposure and the number of lots to
be used will be driven by a manufacturer‟s risk assessment for the IVD and
data from R&D. The risk assessment should take into account at least:
• the variability of the constituent materials (identifying the most important
sources of variability);
• the nature of the users‟ environments; and
• extreme conditions potentially occurring during transportation to those
Users.
Boundary conditions for stability studies should reflect realistic extreme
conditions that are consistent with the design input requirements for the IVD.
The consequent stability studies will prove the IVD capable of meeting
performance requirements at the end of its stated shelf-life, after transport to
the users.
4.4. Minimum number of lots
Similarly to clinical performance validation, the design of stability studies
should take into consideration lot-to-lot variation, with a risk assessment to
identify the most important sources of variability. Lot variability is caused
usually more by the biological reagents than by the actual manufacturing
process. Although existing standards recommend the use of one lot for
certain stability studies, the impact of lot-to-lot variability must be taken into
Page | 9Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
consideration and use of additional lots may be necessary. To ensure the
potential of lot-to-lot variability is addressed, optimally lots containing
different batches of critical constituents such as nitrocellulose membranes,
recombinant antigens, peptides, nucleic acids and enzymes used in nucleic
acid testing (NAT), etc. that are as different as possible should be used.
It is advisable to test in triplicate at each point of testing interval and
maintain product, calibrator and external sample 20% extra quantities or
needs for extra during the study to accommodate all testing need.
Example: For NAT assays, it is critical to use unique enzyme lots for
stability studies. Other components including primer, probe and buffer can
also be affected by the manufacturing process (purity, pH, DNase & RNase
contamination, etc.). For these, different lots are also highly desired that
represent both material and process variability.
4.5. Assessment of liquid components:
It is standard best practice in stability studies to ensure that liquid
components are in contact with all the parts of their container – vial, sachet
or bottle, such as the stopper, the seal and the body of the container. This is
sometimes called “inverted container stability” but is probably best studied
by ensuring all containers are on their sides and disturbed by movement
during the stability study. This aspect needs particular attention for in-use
stability studies of those components that are diluted or reconstituted from
freeze-dried before use.
4.6. Specimens for the stability testing panel:
The specimens used in the stability testing panels must reflect all the
performance claims related to the IVD. If a variety of specimen types (e.g.
serum, plasma, whole blood, saliva) is claimed as being suitable for use in
the instructions for use (IFU), the stability plan must be designed to provide
evidence that the IVD will maintain each of the claims (e.g. sensitivity,
specificity, proportion of valid runs, precision) for each of the specimen
types for the whole of the claimed shelf-life including transport to the final
users.
Evidence should be statistically valid. The stability testing panel must be
validated accordingly and rejection and replacement criteria should be
established. Regulatory requirements may also dictate the addition of panel
members.
A stored validated stability testing panel is not always feasible. For example,
this is often the case for assays requiring fresh and/or whole blood
specimens e.g. CD4, assays to detect RNA. When replacing panel
members, the accuracy of results generated with the replacement material
must be confirmed using an appropriate reference comparator method.
Replacement criteria for unstable panel members will include the duration
for which a critical member will give valid results.
Page | 10Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
4.7. Validation of stability testing panel
The validation of the stability testing panel members used is critical. Stability
testing panel members themselves must be stable, and they must monitor
parameters that are useful to control the component involved.
Stability testing panel members are chosen deliberately to ensure each
member has an attribute pertinent to the intended use. As with lot release
testing, the goal of stability testing is to ensure that the test method
appropriately monitors the functionality of the antigens, epitopes, and
antibodies that are relevant to the intended use at the end of the assigned
(shelf/in use) life.
For instance, the intended use claim may be that early seroconversion
specimens are detected. To show that this claim is true at the end of the
product‟s life, a very early seroconversion specimen is included in the
stability panel. This specimen may be a weakly reactive IgM specimen.
An expected value is then assigned to each panel member and this is used
to assign the acceptance criteria for that panel member. The value for each
member is assigned in a measurable manner relevant to the outputs of the
particular methodology. For instance, the acceptance criteria for each panel
member may be assigned in terms of sample-to-cut-off ratio, cycle time (CT)
values, and band intensity measured semi-quantitatively/quantitatively.
In the example of a weakly reactive IgM seroconversion specimen, the
specimen at the start of shelf life may have a reading score on an RDT of 1+
out of 4, assigned by using a semiquantitative value based on band
intensity. The acceptance criteria may be that all reactive specimens remain
reactive, and all non-reactive specimens do not react in the assay.
As such, panel members must be chosen that not only will be relevant to
demonstrate the intended use, but have values that will appropriately detect
and therefore monitor any deleterious effects of storage. A strong positive
specimen, which has a 4+ out of 4 semi-quantitative reading value, may
remain giving this reading despite decay in the assay, whereas a specimen
with a reading of 1+ out of 4 (with an assigned acceptance criteria of
remaining positive) is more likely to give an indication of the ongoing stability
of the assay.
Thus, it is essential to know that where a panel member meets acceptance
criteria, this is a true reflection of the stability of the product and not due to
the inability of the specimen result to reflect this change.
4.8. Time points
A simple study design requires minimum three testing intervals:
• an initial baseline test and
• a test at the time point beyond the claimed stability limit
• and one point in between.
Page | 11Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
However, this is a high risk approach that has the potential for wastage of
time and resources. If the IVD does not meet the acceptance criteria at the
end of testing there is little information about the deterioration of the
component or IVD (or lack of deterioration) in the interim period.
A more effective approach is to test at predetermined time point intervals.
The manufacturer should decide on practical intermediate test points. The
number and length of testing intervals should be determined in advance and
form part of the stability plan/protocol. This planning will help to understand
the resources required to execute the experiment.
Testing of all panel members is not required at all test/time points. However
testing with all panel members is required at the initial, the second last and
the last test/time point of any of the study specimen types. The
manufacturer should decide on practical intermediate test points at which a
smaller minimal number of panel members are tested. There should be a
documented rationale for the choice of the panels used at the intermediate
test points (e.g. representative members, specimens that are close to the
medical decision points and at the extremes of the assay range tested).
4.9. Duration of testing
Testing conducted in stability studies should extend beyond the shelf-life
determined from the user needs. The shelf-life should be assigned based on
a risk assessment of the lot-to-lot variability in signal change at the end of
shelf life. At a minimum, testing should extend at least one time point (one
testing interval) beyond the determined user requirement. This provides a
safeguard in the event of unexpected IVD failure at the end of the testing
period, in which event extrapolation from an earlier time point would not be
considered acceptable.
It is recommended to utilize standardized units of measure for the entire
study
(e.g. Unopened kit shelf life are always measured in months; opened kit
/reagent stability in days or weeks)
5. Shelf-life studies
5.1. Requirements for determination of shelf life
The stated shelf-life of an IVD must be based on real-time experimental results.
Accelerated studies or extrapolated data from real time data are acceptable for
initial shelf life claim but need to be followed up with real time stability studies.
Such detailed information should describe:
(a) the study report (including the protocol, number of lots, acceptance
criteria and testing intervals);
(b) when accelerated studies have been performed in anticipation of
the real time studies, the method used for accelerated studies;
(c) conclusions and claimed shelf life.
Page | 12Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
Explanation,- Shelf life can be derived from the lot with the longest
real time stability data as long as accelerated or extrapolated data
from all three lots are comparable.
5.1.1. Real-time stability studies
Real-time stability is determined using storage temperatures derived from user
requirements, over a period longer than the required life of the IVD. Where a
range of storage temperature is claimed (e.g. “Store at 4–40°C”), CDSCO
expects the studies will provide evidence for stability over the whole of the
temperature range for at least the length of the claimed shelf-life. Exceptionally,
where claimed stability is restricted to a limited range e.g. “Store at 2-8°C”, it is
acceptable that stability studies are conducted at a single temperature within this
range.
A sequential approach should be used, in which IVDs are first submitted to
stresses simulating transport before they are placed into a shelf-life or in-use
study. This approach best simulates the real-life situation, where products will
first be transported to the end-user and then stored under the recommended
conditions before use, possibly almost until the end of their labelled shelf-life.
5.1.2. Accelerated stability studies
Accelerated stability studies are designed to predict the shelf-life of an IVD from
the increased rates of chemical and/or physical degradation caused by extreme
environmental conditions (e.g. elevated temperature at higher humidity). If the
Arrhenius equation is used to calculate the expected life at temperatures other
than those actually used, then the parameters of the equation must be derived
from the data and not assumed.
Accelerated stability studies provide results in a relatively short time. However
the results of these studies are made using assumptions about the degradation
of reagents and IVD components that may not reflect their performance under
normal conditions of storage and use.
6. Component stability studies
6.1. General principles
6.1.1. Testing on final specifications
Component stability studies, including antimicrobial and desiccant studies,
must be performed using components made according to finalized and
approved manufacturing specifications – ideally to validated
manufacturing scale – on qualified manufacturing equipment and meeting
finalized and approved in- process quality control (QC) specifications.
6.1.2. Considering component stability
Sometimes components of IVDs are prepared in bulk and stored before
being used in several different lots of a completed IVD. The design input
documentation should define how long components are likely to be stored
before use. With that information, component stability studies should be
planned to give evidence that component labelled lives will not restrict IVD
labelled lives: an IVD cannot have a labelled life beyond that of any of its
dependent components.
Page | 13Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
Shelf-lives of components manufactured in bulk and used in several
different lots of an IVD can be verified as for the IVD itself – three lots of
the component as a minimum for shelf-life studies and, depending on
documented risk assessment related to variability, one or more lots
subsequent to change. The evaluated lots of the component must differ in
batches of critical constituents but, again subject to documented risk
assessment, may all be tested in their final presentation with a single set
of the other components which will be used together to constitute the IVD.
Examples of stored components:
Wash solutions and substrates for enzyme immunoassay (EIA),
amplification reagents for nucleic acid testing, calibrators for quantitative
tests, manufactured and stored in their final labelled vials ready to be put
into a kit Component stability can be assessed from the functionality of the
lot and also by factors related to the component itself, such as turbidity,
colour change, microbial contamination and pH of liquid components
changes over time.
Depending on the IVD and the conditions it is subjected to it may be
necessary to distinguish between turbidity that arises from heat/cold
denaturation and turbidity that arises from microbial contamination.
6.1.3. Considering constituent stability
The plan should also consider whether components made from new
constituents (antigens, recombinant antigens, enzymes, antibodies,
membranes) will have the same lives as components made from stored
raw materials. Although this aspect is difficult to study, some evidence
should be provided supporting the use of stored constituents, as well as a
plan to evaluate lot-to-lot variance from different critical constituents.
The choice of the reagents to be used to measure the performance of the
constituent under study (either materials of proven shelf-life or freshly
made) needs substantial consideration.
Examples of stored constituent: Purified recombinant antigens and
monoclonal antibodies stored in aliquots ready for dilution and striping
onto RDT membranes or other supports
6.2. Stability of control materials
Assay specific control materials provided by the manufacturer are to show
that an IVD has performed as intended during use. The manufacturer must
be able to demonstrate that the loss of signal of a control does not occur at
a different rate from the loss of signal from a validated stability testing panel
member or from genuine, critical specimens; otherwise a failed IVD might be
regarded as still functional. Thus the stability of the control material must
accurately reflect the stability of the assay. A control that is more stable than
the IVD and other components, or incorrectly set values for the control
material, must be avoided
Example: It is seen in dossiers relating to IVDs submitted by the
manufacturer that a positive run control will produce a signal of >2.0 optical
Page | 14Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
density (OD) in a freshly manufactured lot, and the IFU will state that an OD
> 0.8 for the same control qualifies a run. Thus, the IVD could have lost
more than half its activity and still appear functional, although some critical
specimens are shown in the dossier to have very weak signals on freshly
made IVDs. This is not considered appropriate unless data can be provided
that demonstrate that the critical specimens will still be detected at the end
of shelf life.
6.3. Antimicrobial stability and efficacy
6.3.1. Rationale
IVDs are used in areas which are not necessarily clean and sterile, and
antimicrobials are not stable under some circumstances. Bacterial and fungal
organisms relevant to the environment of use should be identified in the design
input risk assessment, and antimicrobial preservatives should be chosen to
avoid contamination of the product. The manufacturer must obtain evidence that
the antimicrobial preservative and concentration chosen is stable and effective
against the micro-organisms of concern throughout the claimed shelf-life and in-
use shelf life.
6.3.2. Study conditions
The studies should reflect expected in-use conditions in opened containers:
clean, particle-free laboratories do not usually reflect universal user environment
for suggested methods. Examples of bacterial groups to consider are spore-
forming bacteria, fungus, indigenous bacteria, bacteria found in the environment
of the country of manufacture, as well as use of a negative control. Specific
examples include Aspergillus niger, Bacillus subtilis, Candida albicans,
Escherichia coli, Salmonella species, Pseudomonas aeruginosa, Clostridium
sporogenes and Staphylococcus aureus.
Antimicrobial preservative effectiveness, as measured by the viable microbial
species load present in kit components, should be demonstrated during
development, during scale-up, and throughout the shelf-life.
Testing for antimicrobial preservative content should normally be performed at
release. Under certain circumstances, in-process testing may suffice in lieu of
release testing. The acceptance criteria for in-process testing should remain part
of the specification.
6.4. Desiccant functionality
Desiccants affect the stability of the entire IVD. Stability studies must show that
the desiccant will support the product over the whole claimed shelf-life within the
predetermined extremes of transport, storage and in-use conditions.
Note:
1) CDSCO recommends that a self-indicator (a humidity indicator that changes
colour upon saturation) be part of the desiccant design. However, CDSCO
strongly recommends against the use of cobalt dichloride, the most commonly
used humidity indicator, as it is a carcinogenic substance.
Page | 15Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
2) Sachets are preferable over tablets, since labelling such as “Do not eat” is
more visible. There have been anecdotal reports of desiccants in a tablet
formulation being mistaken for antimalarial medicine.
7. Stability during transport
7.1. Rationale
Transport stability studies evaluate the tolerance of an IVD to the kinds of
environmental conditions (e.g. temperature, humidity, dust) and physical
conditions (inversion, vibration, physical handling, stacking) to which it is likely to
be subjected in the time between shipping from the manufacturer to its final
user. They should provide evidence that there will be no impact on the IVD
performance over the whole of its stated shelf-life after recommended
transportation methods for the IVD. The manufacturer should assess the
potential impact of multiple factors and justify and document whether or not to
include them in the evaluation.
CDSCO expects that a transportation challenge should precede the real-time
determination of shelf-life This serves to determine that transportation conditions
do not reduce the shelf-life of the IVD.
In some cases, it might be acceptable to test the product only over the transport
simulation duration, without a subsequent long-term study under normal storage
conditions. If that is done, shelf-life must be established under specified storage
conditions along with a stringent, evidence-based risk assessment of the
probabilities of extreme transport stress affecting the performance at the end of
the claimed life.
7.2. Challenge conditions
Determination of the stability during transportation of an IVD should take into
consideration the local routes, transport means and transit used to supply the
IVD, usually defined in the design input risk assessment. It is not necessary to
test the IVDs to the point where it is no longer usable, but merely to validate the
window of transport conditions within which the IVD will retain its claimed
performance to the end of its stated shelf-life. However, knowledge of the
possible limitations of an IVD and at what point the IVD becomes unusable is
useful to a manufacturer when trouble shooting post-market problems. CDSCO
expects the manufacturer to consider that the product might continue to be
subjected to sub-optimal storage conditions at the end-user.
Example: While a static challenge of 45°C for 3 days might represent conditions
seen during actual transport of a IVD, a more stringent challenge of cyclical
higher and low temperatures (including freezing) for a longer period of time and
under vibration might better cover a „worst case scenario‟ of shipment, storage
and subsequent transportation to the end-user.
7.3. Number of lots
For transport studies, at least one lot of the IVD can be used.
Page | 16Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
7.4. Multiple stress test sequences
Appropriate sequences may be developed on the basis of data from actual
product transport studies. Testing multiple stress sequences allow a
manufacturer to identify the most cost- and/or resource effective transport
conditions from a set of alternatives while ensuring adequate product
stability protection.
Note: Environmental conditions investigated as part of a stability study must
reflect those likely to be encountered in resource- limited States.
Temperatures at some airport tarmacs can exceed 40°C while temperatures
encountered during air transport fall below 0°C. Significant delays can be
encountered at all times and especially during wet season transport to
remote health centres.
7.5. Physical conditions
Physical handing can be both manual and mechanical. The relevant user
and commercial factors should be identified as part of the design input risk
assessment and the packaging and shipping methods developed
accordingly. Reference defines a number of factors to be considered, and
their evaluation: drop, impact, compression, vibration, repetitive shock,
longitudinal shock, cyclic exposure, vacuum, impact, inversion; along with
the size, weight, and composition of the packaging.
7.6. Simulated versus actual challenge
An actual shipping challenge can be used to verify the conditions found in
the simulated transportation challenges. However, it should only replace a
simulated shipping challenge when there is an appropriate risk evaluation
and with experience and data already actively collected from similar
products and documented in detail (for example it is insufficient to note “no
complaints”). In the R&D phase, actual data from shipping can be used to
define the conditions needed for an appropriate simulation of extremes.
However, in the post-production phase actual shipping challenges often do
not explore the full range of shipping conditions that could be encountered,
including extreme values.
8. In-use stability studies
8.1. Rationale
In-use stability of an IVD is the period of time over which components retain
adequate performance, after transport to the users, once they are opened,
reconstituted and/or diluted and exposed to the environmental conditions in
which they will be used.
If a range of conditions for use is stated in the IFU (e.g. use at 15–40°C)
evidence should be provided to prove the stability over that range with all
the specimen types (e.g. serum, whole blood, oral fluid) claimed. It is
considered best practice that the manufacturer extends the stability range
by 5°C at the lower and upper end of the proposed acceptable range on the
Page | 17Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
labelling for all components to ensure that that the claimed stability ranges is
acceptable.
8.2. Conditions of use
Determination of the in-use stability of an IVD and/or its components should
reflect routine conditions of use of the IVD. Freeze-thaw stability should be
considered to address reagents which are exposed to multiple freeze-thaw
during use.
Note: In-use stability studies must take into account environmental
conditions and usage conditions encountered by users and States, such as
exposure to extreme temperatures, humidity, dust, light and micro-
organisms.
8.3. Multiple in-use stability claims
Depending on the way in which the IVD is used it may be necessary to have
several in-use stability claims. In situations where multiple stability claims
are made, a manufacturer must provide evidence from testing that
investigates routine use supporting each of the claims.
Examples:
1) A reagent may have a stated period of stability once it has been placed
on-board an instrument and another period of stability once it is in active use
(i.e. during actual use/testing).
2) Multiple use reagents (e.g. buffers) may repeatedly be exposed to high
temperatures during the day while in use and exposed to lower
temperatures when not in use and stored in the refrigerator. The actual use
of the multiple use reagent – squeezing of bottles, exposure of the lid and tip
to working surfaces, hands, exposure to dust and light – also affect stability.
Stability studies should take into account all of these factors.
9. Lots used in stability studies
9.1. Considering variability
Stability studies must take into consideration all possible sources of
variation within and between manufactured lots. For most IVDs it is likely
that differences between batches of the biological reagents will cause the
most variance. Factors to consider include apparently minor, technically-
uncontrollable differences in culture and purification for recombinant
antigens and antibodies; synthesis and purification for primers,
probes and peptides; undocumented production changes of an outsourced
buffer component and the lot of nitrocellulose membrane used in lateral-flow
IVDs. At a minimum, lots chosen for stability studies should be different in
the critical constituents, e.g. different purification and/or culture batches for
all recombinant antigens and monoclonal antibodies. If pilot or small scale
lots are chosen, special attention must be paid to the potential for variability.
However, the sources of variation will depend on the particular process,
product and component, and should be identified during product
development risk analyses.
Page | 18Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
Use of different batches of critical components ensures that the stability
evidence obtained is more likely to be representative of long-term
manufacture. Any variability found can be taken into consideration when
assessing the outcome of the studies against the design input requirements
and when making claims. This minimizes user problems and hence
complaints.
9.2. Testing the final configuration
Shelf-life, in-use and transport stability must be determined for the finalized
product configuration, in terms of:
• manufacturing specifications;
• release-to-market QA criteria;
• packaging and labelling; and
• validated manufacturing scale on qualified manufacturing equipment.
Note:
Testing methods should be as included in the IFU of the finalized IVD.
It is important that it can be established that the stability studies were
conducted on the IVD as submitted to CDSCO for approval. Even changes
perceived as small (e.g. change in production scale, bulk container
materials, supplier of a critical biological, change in vial stopper) can have
unexpected effects on stability and other performance characteristics. After
such changes, a stability plan and study is needed again. Manufacturers
should have change control procedures in place compliant with ISO 13485.
Stability studies undertaken in the R&D phase of the product lifecycle are
important to understand how to design the product so it will meet the final
stability requirements in the input documentation. However, these studies
are not sufficient for submission to CDSCO since they might not reflect the
final design and manufacture of the IVD.
9.2.1. Exceptions
If any of the above criteria are not met (for example if “pilot lots” or small
scale lots are used, or if the IFU is not finalized), strong evidence must be
provided that the evaluated materials will perform exactly the same as the
final product.
Note: In some exceptional circumstances, where it is not possible to sample
from actual production lots, samples from pre-production or development
lots might be used. If this is the case, manufacturers should justify why
production lots were not used, and they should provide robust evidence that
the lots chosen are expected to behave identically to the production lots.
Data concerning lot-to-lot variability must still be submitted. Although
CDSCO will consider the available evidence on its merits, this preliminary
information must be followed by stability claims conducted on production
lots. A post-Approval commitment may be required to amend this situation
when the manufacturer is able to produce fully qualified production lots.
9.3. Number of lots required for testing
Page | 19Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
As per MDR 2017 Testing should be performed on at least three different lots
manufactured under conditions that are essentially equivalent to routine
production conditions (these lots do not need to be consecutive lots) be used
to verify shelf-life; one lot reflecting actual routine use of the device (real or
simulated) be used to verify in-use claims; one lot to evaluate the tolerance
of products to the anticipated shipping conditions be used to verify Shipping
stability.
Note:
It is not acceptable to sample IVDs from a single manufactured lot but label
them so that they appear to have been taken from three separately
manufactured production lots. This aspect will be investigated during an
onsite inspection by CDSCO. Non-compliance with this requirement may
result in a major Non-compliance under the MDR 2017.
9.4. Components of lots required for testing
Stability work is performed using materials in their final packaging, with
intended labelling. If there is more than one variant of the IVD (e.g. pack
size differences) any potential effects on performance, including stability
must be assessed.
In particular, if different reagent-container sizes are used in packs intended
for different numbers of use, stability evidence should be obtained on all
variants, even if the contents of the containers are identical.
Once component shelf-lives are assigned use relatively fresh components
and components which have progressed into their assigned shelf-life in the
different production lots used in the establishment of the product shelf-life.
10. Stability protocol
Stability studies should be well designed, scientifically sound, well
implemented, well recorded and able to deliver meaningful conclusions about
IVD performance.
This will minimize the time and resources taken by the manufacturer to
generate appropriate evidence and by the regulatory authority to assess it.
It is good practice to prepare, within the mechanisms of a quality management
system (QMS), a plan for the investigation of each aspect of IVD stability. A
well-developed study protocol, with clearly defined objectives, responsibilities
and pass/fail criteria should be developed, reviewed and internally approved in
advance of testing. The protocol should be associated with the design input
requirements.
It is essential that the study protocol takes into account the intended use of the
product to ensure that these elements are covered within the stability studies.
The results of the stability studies support the claims in the instructions for
use. Careful forward planning will make a significant contribution to ensuring
that sufficient resources are made available, effective experiments are
performed and both experimental results and associated documentation are
recorded in an appropriate manner.
Page | 20Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
10.1. Responsibilities
The study protocol should outline responsibilities and applicable training for
said responsibilities of all staff involved in the study. The R&D department is
usually responsible for set up of the study and testing of newly developed
IVDs, monitoring, and any equipment validation if required, and for the
documentation of the testing plan and sample selection.
The R&D department should nominate a responsible person for investigating
failures. The QA department should nominate a responsible person for
conducting risk assessments, if IVD fails to meet the requirements of the
design inputs. Performance evaluation is not to characterize an IVD but to
show that it meets (or exceeds) predetermined qualities.
10.2. Preparing the testing plan
A complete, detailed description should be prepared that fully documents
everything to be done and the expected outcomes. Authorization of the
protocol should be obtained internally in advance of starting work. The
protocol should include the following details.
• Qualification and training of technical staff performing the work
• Biohazard issues identified with reagents
• The instrumentation, including storage facilities or rooms, validation,
calibration, monitoring, servicing
• The batch numbers of kits to be used with justification for any
manufacturing anomalies or excursions from documented procedures
• The expected life of the kit from the input documentation
• Any proposal, with justification to launch a kit with a life based on
accelerated data, or to launch with a shorter life than in the input
documentation while awaiting the conclusion of real-time testing
documented.
• The documented nature and extent of in-use testing
• The justification for the choice of lots and components taking into account
lot-to-lot variation and the critical characteristics
• The number of units (cassettes, bottles, tablets, etc.) of each component
to be collected and stored under each condition
• The nature of the stability testing panel to be used, justifying each panel
member‟s inclusion and defining the volume and characterization of the bulk
specimen to be used and the aliquot size and number to be stored for the
testing
• The expected criteria for each stability testing panel member at the
beginning and end of the product‟s proposed shelf-life
• The statistical methods to be used for data analysis
• Graphs (paper or electronic) to visualize the performance of each stability
testing panel member over the course of testing
• Methods of approval and justification of any deviations from the plan
10.3. Product storage
Page | 21Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
A sufficient number of product components from the identified lots should be
reserved and stored separately to ensure that the study will be completed
with identified products. Sufficient volumes should be retained to allow for
the predetermined invalid rate.
10.4. Documentation
The Stability protocol should make reference to a study report which will be
used to summarize interim, and ultimately, final study findings and
conclusions. The study plan, the testing protocol and study report and all
associated documentation (worksheets, etc.) should be controlled within the
manufacturer‟s QMS. At the end of the study, the manufacturer should be
able to confirm that design input requirements have been met.
Any changes in method must be recorded and undergo risk assessment. It
should refer to the development of a detailed and valid testing protocol
which includes all information and material relevant to testing.
10.5. Statistical methods
Statistical methods are used to support stability claims by providing
estimates of the probability of results being as stated. For example: prior to
the stability studies on an EIA it has been documented that if a stability
testing panel member has at least a particular optical density (OD) then that
device will meet a particular claim. Given the results of the stability study
using that stability testing panel member and showing the variability within
and between lots of the IVD, the probability of future similar production of
the device meeting claims at the assigned life can be estimated. The
derivation of valid criteria and the probability of maintenance of all claims
can be estimated by appropriate statistical methods.
A fundamental problem is that of how many replicates should be used at
each time point and from how many different production lots to produce
acceptable overall probability estimates of the likelihood of all future
production of similar devices and lots meeting claims (and hence user input
requirements) at the end of the assigned life. There are two aspects to this –
what is “acceptable” and “how many replicates?” “Acceptability” is a decision
critical to quality and must be decided in advance from the user
requirements – for example 80% confidence that 95% of all lots will meet
the claims. This is in fact a tolerance interval as described in ISO 16269-
6:2014. “How many replicates” can then be derived from the tolerance
interval required but advice from a professional statistician is strongly
advised – after defining the quality critical requirement but before beginning
any experimental work.
The statistical methods to be used will be documented in the plans and
protocols of any stability study and consideration given to treatment of
unexpected and atypical results. In general all results must be used unless
there is a documented physical reason (e.g. known operator error, too little
volume, incorrect timing, use of an unqualified instrument such as lacking
Page | 22Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
maintenance or calibration) why a result can be ignored – but even then that
result must be recorded and included in the report of the stability work.
10.6. Stability testing protocol
As part of an approved study plan for the determination of IVD stability, a
detailed testing protocol should be prepared (examples of stability protocols
are provided in Appendix 1: Example stability protocols) including the
following as a minimum, as appropriate.
QMS identifiers (e.g. experiment name, document references, etc.)
that allow traceability to both the overarching study plan and to
subsequently generated records/documents such as result
worksheets
The name(s) of operator(s)
The dates and times when the experiment was performed
Signatures of the operator and supervisor(s)
The objectives of the study (i.e. determination of shelf-life,
determination of in-use stability of a component, etc.)
The name and lot number of the IVD and/or components being
investigated
How the components will be sampled from the production department
Stability testing panel members and their characterization to be used,
including valid test methods which reflect the IFU claims
The experimental method that will be used for testing. This must
follow the finalized testing method from the IFU. It must describe
clearly how the experiment was performed in terms of:
required storage and/or challenge conditions;
the duration of storage/challenge;
the schedule of testing intervals;
the stability testing panel; and
the numbers of replicate tests performed for each stability testing
panel member.
How and where results are to be recorded
Acceptance criteria
How aberrant, discordant or invalid results will be dealt with
How storage/challenge conditions are to be applied
Example: For determination of stability during transportation it should
be made clear that each IVD will be subjected to a sequence of stated
temperatures.
How actual storage/challenge conditions are recorded
Example: Recording of temperature not as “room temperature” but as
an actual numerical value obtained from calibrated instrumentation
Note: It can be unclear to a CDSCO reviewer from a general statement
such as “… Sample buffer was stored at the required temperature and
tested each month…” whether (1) the bottles of sample buffer were stored
Page | 23Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
open at the required temperature for the entire testing period, or (2) the
bottles were stored capped and refrigerated, and only reopened briefly at
the required temperature at each schedule test point.
10.7. Reading and recording results
10.7.1. Avoiding reader bias
It is good practice to use approaches to make the reading more objective,
such as a scoring system. For IVDs where a subjective element forms part
of the result, e.g. reading the intensity of an RDT band within a specified
time frame, the results should always be reviewed by a first and second
reader to avoid operator bias. Both readers must be blinded to the
expected results; the second reader must be blinded to the first reader‟s
results. If a validated band intensity scoring tool is to be included in the
final RDT kit, this should be used to record results.
10.7.2. Recording actual individual results
The results of a test, not only the test interpretation, should be recorded.
An interpretation on its own has insufficient resolving power to allow
degradation of a signal over time to be observed. Some IVDs, e.g. line-
blots, may require particular band patterns to allow an interpretation to be
reached, and several different patterns may yield the same final result.
Recording only the final interpretation of a test specimen may cause the
failure of particular bands to go unnoticed while allowing the IVD to
otherwise “pass”. Photographic records of qualitative tests are
recommended, as appropriate.
This is particularly important when testing a panel of like specimens, e.g.
“20 HIV antibody positive specimens” for which the acceptance criterion is
“all 20 specimens must be positive”. It is not sufficient to simply record “all
20 positive” or “pass” without first recording the individual test result
directly from the IVD for each specimen in the panel.
Example 1: For most enzyme-linked immunoassays (EIAs) if the sample-
to-cut-off ratio is > 1 then the result is interpreted as “positive” or
“reactive”. In this case three pieces of information should be recorded: (1)
the numerical value of the assay sample-to-cut-off ratio, (2) the numerical
value of the signal for the specimen and (3) the final interpretation.
Example 2: Some rapid diagnostic tests (RDTs) may stipulate that the
strength of test band is not correlated with the strength of antibody titre.
Nevertheless, the following should be recorded: (1) the intensity of
observed patterns according to a predetermined, validated intensity
scoring system with as fine a gradation as possible, and (2) the final result
interpretation.
Example 3: A qualitative NAT assay may report “positive” and “negative”
for a particular analyte, but the underlying decisional parameter is often
quantitative (e.g., a PCR signal-based cycle number). The quantitative
parameter should be recorded.
Page | 24Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
10.7.3. Retention of records
CDSCO encourages retention of photographic records, machine printouts,
electronic data or physical retention of membranes from opened
cassettes, as appropriate. Records should be retained for a period of time
at least one year after the date of expiry of IVD as defined by the
manufacturer, but not less than two years from the date of product release
by the manufacturer
10.8. Degradation vs deterioration
Testing at more than two time points can be important to avoid confusion
between imprecision and stability. For example, if the end testing shows
10% decrease, one may not judge if the difference was due to imprecision
or degradation. If tested one or more times in between are used, fluctuation
caused by imprecision can be distinguished from drift due to instability. This
can be ameliorated by increasing the number of replicates and runs. All
studies should support precisely defined periods of in-use stability claims.
Example: An RDT test cassette – may be labelled “Use immediately on
opening”. In such cases it is still necessary to determine the interval (one
hour, one day, etc.) over which the IVD performance remains stable after
the component is opened.
10.9. Testing schedule
Testing intervals should be selected to detect any trending activity over the
testing period. Concurrent testing of separate types of components may be
approached with different intervals. For example, it may be appropriate to
test an IVD test cassette against a stability testing panel on a monthly or
quarterly basis.
10.9.1. Acceptance criteria for results
The acceptance criteria to establish what is acceptable or not acceptable
should be defined according to the stability testing panel criteria for both
qualitative and quantitative test methods. Results from failed (invalid) test
runs must not be used in the determination of the stability claim. However
the invalid results should also be recorded.
11. Stability report
11.1. General
After testing has been completed, the findings should be summarized in a
stability study report. The report should clearly identify the IVD that was
tested, the objectives of the study, the conditions under which the IVD was
tested and conclusions that were drawn from findings. The report should be
traceable to the study plan, testing protocol and user needs. It should make
clear references to other supporting documentation (e.g. result worksheets).
11.2. Link to claims
Page | 25Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
The results and conclusions of stability studies presented in the study report
must support the claims of IVD stability reported in the IFU and elsewhere in
the dossier.
11.3. Consider variability
An overall stability claim (whether for shelf-life, in-use stability, or stability
during transportation) must be based on the expected stability when taking
into account inter-lot variability.
Example: The manufacturer should evaluate the variability between the
different lots studied and assume that any differences in shelf-life are
inherent to the manufacturing process. The claimed life should be calculated
so that a known and stated proportion of all lots (usually >95%) will meet the
claimed shelf-life. Frequently more than three lots are needed to obtain a
realistic idea of the variability of the results.
11.4. IVD stability versus component stability
A claim of stability for an IVD as a whole must not exceed any individual
component stability.
Example: For an IVD claimed to detect HIV-1 and HIV-2 antibodies – if
detection of HIV-1 antibodies is stable to 24 months but that of HIV-2 to only
18 months, then the shelf-life must be based on the shorter time.
12. Changes to a Licensed / Approved IVD
Any major modification to a licensed / approved IVD or to its process of
manufacturing will require provision of direct evidence of stability. An
appropriate risk analysis and an accelerated stability study comparing the
original product and the modified product for usability, performance and lot-to-
lot variation may serve to assess the impact of the changes to a product
formulation or manufacture. It would be necessary to validate the stability of
the modified IVD in at least one lot of the IVD (subject to risk analysis) in order
to demonstrate equivalence between the original and modified IVDs. More lots
may be appropriate depending on the product nature, variability of
components and failure risk. CDSCO expects results of accelerated testing to
be confirmed by real-time studies.
If there are different presentations, the stability of each one must be assured.
The following examples seek to illustrate the scope for considering the
performance evidence from one IVD as support for performance in another:
Examples:
1) For an HIV RDT which uses an identical cassette and physical components
of a manufacturer‟s existing, fully validated HCV RDT, the reagent
formulations are different (antigen/antibodies, buffers, conjugates, etc.).
Evidence of stability of the HCV RDT would not suffice for the HIV RDT. Even
if the manufacturer claims that both IVDs have been sold in a number of
countries for several years and no adverse feedback has been reported, this
would not constitute evidence in support of the stability of either IVD.
Page | 26Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
2) For an HIV RDT that has been fully validated for detection of HIV-1
antibodies; a new product is developed which includes detection of HIV-2
antibodies. The stability of any sample buffers that are identical between the
two IVDs would probably not need to be validated. However, other
components (conjugates, antigens, antibodies) that are different between the
two IVDs would need to be tested; it would not be sufficient to assume that
HIV-1 reagents will have the same stability in the new IVD. A modification of
this nature is likely to require substantial validation of stability.
3) An HIV RDT IVD previously intended for testing serum/plasma has added
to it a claim for detection of HIV-1 in whole blood. The only substantive design
change associated with the new claim is the addition of a small pad of some
suitable material near the sample port which acts as filter for whole blood
specimens.
Depending on the nature of the material it may be reasonable to argue that the
material would not be expected to age; that it is not, in any practical sense,
chemically labile. Consequently, shelf-life and in-use stability may not
necessarily need to be retested in full. However, stability during transportation
may need to be determined to provide confidence that the modification is able
to withstand likely shipping conditions (e.g. that the extra square of filter paper
doesn‟t dislodge when packages are jostled and bumped in transit).
4) Based on an HIV RDT that has been fully validated for detection of HIV-1
antibodies, a new IVD is developed which includes detection of antibodies to
Treponema pallidum (TP). Detection of TP specific antibodies occurs on a
completely separate membrane (and associated architecture) to that of HIV
antibody detection. Additional handling steps may have an impact on the
stability of the HIV-1 antibodies and it may be required to retest. It may be
necessary to review evidence of stability during transportation to ensure that
new components are not affected by transport (for example a new packaging
concept is used).
If a new machine is used for striping of the HIV-1/TP IVD, validation of the new
machine (installation qualification, operational qualification and performance
qualification) would be required to show that the stability studies are still valid.
If the IVD is designed in a way that HIV and TP detection occurs either on the
same membrane and/or using most of the same architecture (and assuming
that sample buffers are identical between IVDs) it is likely that this new IVD
would need to be fully validated.
It should be noted that these observations pertain specifically to IVD stability.
Other aspects of IVD performance should still be validated as appropriate.
Page | 27Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
APPENDIX
Annexure I Example stability protocols:
Annexure II Suggested specimens for stability testing panels
Annexure III Examples of Stability Study approaches
Page | 28Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
APPENDIX – I
Example stability protocols:
This appendix contains examples for a wholly fictitious IVD, illustrating the kinds of
experimental design to determine the following:
1. Stability of whole kit during transport
2. Stability of whole kits during shelf-life, and
3. In-use stability of whole kits including reagents
The information provided in these examples should not be taken as a checklist of
sufficient conditions, but manufacturers are encouraged use as a guide on possible
approaches to generate evidence of a standard sufficient to satisfy the equirements
of the MDR2017. It is recommended that transportation stress studies are
undertaken prior to the shelf-life studies.
Description of fictitious IVD
IVD used for the purpose of the examples is a RDT for the detection of antibodies to
HIV-1, HIV-2 and Treponema pallidum in serum, plasma and whole blood. It is
recommended that the kit is stored at 8–40°C, but components of the kit must be
used at 15–30°C. The product is supplied as a kit with each test cassette sealed in
a foil pouch (with desiccant). The pouch must be brought to 15– 30°C. Once
opened, it is recommended that the cassette is used immediately. The IVD includes
a bottle of specimen buffer/diluent for use with all three specimen types. The
specimen buffer is expected to have similar stability as the test cassette in its
unopened form. The stability of the opened bottle of specimen buffer is determined
below (see Example 3: In-use stability protocol).
The manufacturer of this product proposes to determine the stability of its product
and has written a stability plan. As part of this plan a preliminary determination of
accelerated stability has been conducted at several extremes of temperature and
suggests that the IVD would be stable to an equivalent of 12 months following
manufacture. The plan now calls for the development of real-time stability protocols
that will form the basis of subsequent testing of the IVD.
Preliminary work has shown that the variability between lots is minimal so that three
independent lots (no critical constituents in common) will suffice to enable a
reasonable estimation of shelf-life taking lot variation into account.
Page | 29Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
Example 1: Evaluation of stability during transportation
Objective:
To determine the stability during transportation of the HIV RDT in real-time using
simulated shipping conditions and to generate stressed components to be used in
real-time shelf-life studies as proposed in Stability Study Plan xxxxx01.
Preparation
Acquire sufficient numbers of kits from three independent production lots using a
Predetermined sampling protocol (e.g. random, first X kits in first box, every 100 th
kit, etc.). Allow at least 10% for unexpected requirements and re-testing
Note 1: To provide security against unforeseen events, duplicate tests should be
performed as a minimum. Testing in triplicate as a minimum provides a level of
statistical confidence in the observed test result.
Testing will be conducted at 0, 3, 6, 9, 12 and 13 months.
Note 2: Testing beyond 13 months will allow an understanding of when, in real-
time, the IVD is likely to „fail‟ and may allow an extension of the proposed shelf-life.
Note 3: For determination of shelf-life a fresh bottle of specimen buffer must be
opened at each testing point – although there may be circumstances in which
multiple sampling could be taken from the same bottle after it has been opened.
Acquire sufficient volume of each stability testing panel member for the duration
of the testing schedule.
The protocol for these studies specifies the number of devices to be picked, the
statistical sampling plan to be used and the required stability testing panel
members and their volumes.
Documentation:
In Worksheet xxxxx01 record the following:
• The lot numbers from which kits were sampled
• The number of kits sampled from each lot
• Details (including manufacturing/lot information) for each of the kit
components that will be tested as part of this protocol:
Test cassette:Bottle of Sample Buffer:…
The product kits chosen to be tested are in their final packaging including all
labelling.
The IVDs are stored so that the reagents are in contact with all elements of the
Page | 30Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
packaging (e.g. the bottles in the product kits are stored horizontal lying flat on
their sides).
Kits will be divided into two groups. One group will be stored at 40 ± 5°C , the other
at 8 ± 2°C . Kits from each group will then be subjected to the following conditions.
Testing schedule: for transport simulation
Condition 1, Temperature and humidity sequence: all kits will be taken through a
temperature and humidity sequence consisting of:
i) Ambient humidity (X% RH)
Put at IFU storage temperature for 24±4 hours followed by
30 ± 5°C for 24±4 hours followed by
45 ± 5°C for 24±4 hours, followed by
8 ± 5°C for 24±4 hours, followed by
IFU storage temperature for 24±4 hours
Followed by
ii) Desert humidity (30% RH)
Put at IFU storage temperature for 24±4 hours followed by
30 ± 5°C for 24±4 hours, followed by
45 ± 5°C for 24±4 hours, followed by
8 ± 5°C for 24±4 hours, followed by
IFU storage temperature for 24±4 hours
Followed by
iii) Tropical humidity (85% RH)
Put at IFU storage temperature for 24±4 hours followed by
30 ± 5°C for 24±4 hours, followed by
45 ± 5°C for 72±4 hours, followed by
8 ± 5°C for 24±4 hours, followed by
IFU storage temperature for 24±4 hours
Followed by
iv) Ambient humidity (X% RH)
Put at IFU storage temperature for 24±4 hours followed by
30 ± 5°C for 24±4 hours, followed by
45 ± 5°C for 24±4 hours, followed by
8 ± 5°C for 24±4 hours, followed by
IFU storage temperature for 24±4 hours
Note 1: It is important to make clear that the above complete sequence of
temperatures will be used, as opposed to separate IVD kits being held at individual
temperatures. The actual temperatures, durations and the nature of the sequence
Page | 31Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
will depend on the IVD and the kinds of conditions expected to be encountered
during shipping
Note 2: Freezing temperatures are not considered in this example but should be
included if the IVD kits could be exposed to freezing temperatures during transport.
Note 3: If transport by air is anticipated, the effect of reduced pressure should be
included in the protocol (3) for a period of time at least 10% longer than the longest
anticipated flight, and at a pressure expected in aircraft holds.
Note 4: The protocol should call for testing of at least five individual IVD kits after
each stress condition, using the stability panel members giving the most informative
results. This approach will enable verification that the IVD kits are sufficiently stable
to progress to the next condition, although this should already be known from
preliminary experiments and R&D work.
Condition 2, Transport stress conditions - Shaking. Each IVD kit will be placed on a
shaking table at X revolutions per minute (rpm) for X hours/days at 42 ± 5°C as
defined by ASTM D4169 section 12 (3).
After the simulated shipping challenge, each IVD kit will be returned to its
corresponding storage temperature (42±5°C or 8±2°C).
Testing schedule for real time stability studies
Testing will be conducted at 0, 3, 6, 9, 12 and 13 months. At each scheduled time
point, the allotted number of IVD kits will be brought to 15 to 30 °C and used to test
each member of the panel in triplicate.
Note 1: The test at 0 months will provide evidence that the IVD kit is stable under
extreme conditions of shipping (but similar to those likely to be experienced), the
testing at later time points will give evidence to support the claimed shelf life after
transport, and testing beyond the claimed shelf life will provide evidence that the
IVD kit is stable and not close to a failure point.
Documentation for transport stress conditions
In Worksheet XYZ00001 record:
The lot numbers of the IVD kits used to conduct the test
The Operator(s) name(s)
The dates of testing
Identifying details for each member of the panel being tested
The temperature at which the IVD kits are stored
The values of temperature and humidity for each of the challenge conditions
Instrument settings for the shaking apparatus and duration of operation for
challenge conditions.
The ambient temperature and humidity during testing
Each test result as an interpretation according to the IFU
Each test result as a band intensity. Band intensity should be scored using
the calibrated scale described in Protocol ZXY0001 (e.g. 0, faint/trace, +1,
+2, +3 … +10) (even though the IFU does not give scores to results)
Page | 32Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
Any aberrations or deviations from the protocol, the reason for the deviation
and any remedial action undertaken. Results from invalid assays must be
recorded but not included in calculations of shelf life. Apparently aberrant
results, unless the underlying cause can be positively identified as not
related to a problem with the IVD, must be included in the calculations of
shelf life.
Acceptance criteria
Each panel member should show a band intensity result that matches its expected
result at each tested time point. The expected result must be validated so that if the
IVD fails to meet the claims (e.g. fails to detect critical specimens, has unacceptable
performance at medical decision concentrations, has unacceptable specificity) the
panel member would also fail to meet its specified result.
The stability after transportation of the IVD kit will be taken as the time point before
the last time point to have met the acceptance criteria, e.g. if the IVD is stable to 13
months, the stability after transportation will be deemed to be 12 months.
The stability after transportation should be identical to the claimed shelf life of the
IVD kit, i.e. the extremes of possible conditions to which the IVD kit is likely to
subjected during transport must not affect the shelf life of the IVD.
Calculation of results
Detailed statistical instruction must be obtained from a professional statistician with
an understanding of the expectations of the stability study plan and outcome.
Professional statistical input is particularly recommended when calculating
confidence limits for discrete data such as readings from a graduated scale.
Each of the following applies at each time point:
The variance of the results for all replicates within and between all the lots must be
calculated for each panel member. From the overall variance between lots, the
confidence with which future lots of the IVD kit will detect the panel member at that
time point after manufacture and transport can be calculated. If the confidence of
the panel member meeting its specification is less than some pre-defined value
(normally 95%), it must be deemed to have failed at that time point and the shelf life
of the IVD kit should be restricted accordingly.
If regression analysis is used to define the time point at which a panel member
would not meet its criterion, then lot-to-lot variability must be included when setting
the confidence limits around the regression line. However, real-time data must
extend beyond the claimed shelf life so that the intercept of the regression
confidence limit and the expected value must be at a time period longer than the
claim. It is usually more appropriate to calculate as discussed in the previous
paragraph, particularly if the regression cannot be proven to be linear.
Page | 33Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
Example 2: In-use stability protocol
Objective
To determine the stability of opened bottles of the Specimen Buffer used in the IVD
kit in real-time when stored at 15–30°C as proposed in Stability Study Plan
XYZ00001.
In this example the manufacturer recommends that the test cassette be used
immediately upon opening; this claim should also be validated in a separate
experiment, so that it can be confirmed that the IVD will still perform satisfactorily
after the test cassette has been removed from its pouch and open at room
temperature for 1, 2, 6, 24 hours, etc., as appropriate.
Acquire sufficient numbers of IVD kits from one production lot using a
predetermined sampling protocol (e.g. random, first X number of kits in the first box,
every 100th kit, etc.).
Acquire sufficient volume of each panel member for the duration of the testing
schedule. Establish a method for randomising the panel for testing.
In Worksheet XYZ00001 record the following:
The lot numbers from which the IVD kits were sampled
The number of IVD kits sampled from each lot
Details (including manufacturing/lot information) for each of the IVD kit
components that will be tested as part of this protocol (test cassette and
specimen buffer).
Preparation
Two lots of specimen buffer are to be tested. One lot of the component must be
freshly made, the other should be towards the end of the assigned shelf life of the
IVD kit.
The component is to be tested in its final packaging.
The IVD kits are stored so that the reagents are in contact with all elements of the
packaging (e.g. the bottles in the IVD kits are stored horizontally, lying flat on their
sides, allowing liquids to remain in contact with the bottle closures).
Page | 34Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
Half of each lot will be stored at 30± 5°C, the other half at 15± 5°C. At the start of
testing each bottle will be brought to room temperature (20 ± 2°C), opened, used for
testing and then recapped and returned to the stated storage temperature.
Note 1: It is important that the components under test are opened and used under
circumstances likely to occur in users‟ laboratories (i.e. not in rooms with HEPA
filtered air) mimicking, as far as possible, genuine use.
Testing schedule
At each subsequent scheduled time point the allotted number of bottles will be
brought to room temperature and used to test each panel member in triplicate.
Testing will be conducted at 0, 1, 2, 3, 4 weeks, etc., up to the end of the claimed
in-use life.
Documentation
In Worksheet XYZ00001 record:
The lot number of the IVD kit used to conduct the test
The Operator(s) name(s)
The dates of testing
The temperature at which the IVD kits are stored
The ambient temperature during testing
Identifying details for each member of the panel being tested
Each test result as a band intensity. Band intensity should be scored using
the calibrated scale described in Protocol ZXY0001 (e.g. 0, faint/trace, +1,
+2, +3 … +10)
Each test result as an interpretation according to the IFU
Any aberrations or deviations from the protocol, the reason for the deviation
and any remedial action undertaken.
Acceptance Criteria
Each panel member should show a band intensity result that matches its expected
result at each tested time point. The in-use stability of the sample buffer will be
taken as the time point before the last time point to have met the acceptance
criteria.
Example: If the IVD kit is observed to be stable to 5 weeks, the in-use stability will
be deemed to be 4 weeks.
APPENDIX – II
Suggested specimens for stability testing panels
Page | 35Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
Examples in this section
Not all of the specimens in the examples that follow will be necessary for all IVDs,
nor is the list exhaustive. Panels must be composed according to strict risk
management principles, and all decisions must be documented and traceable.
The minimum specimens that are recommended to be included in a testing panel
for the different products are outlined below.
1 Specimens to monitor tests for nucleic acid-based testing technology
If a proprietary nucleic acid preparation /extraction system is provided, the recovery
must be shown to meet claims for each genotype from each of the specimen types
claimed (e.g. dried blood spots, whole blood, plasma). Successful removal of
inhibitory substances, if intended, must be demonstrated for appropriate specimen
types. Unless potentially variable biological reagents are involved, this system
would be expected to be verified in manufacture and not necessarily tested at
release.
Specimens Remarks
Specimens to Traceability is required to one of the standard of NIB or
demonstrate NIBSC or WHO or any other, which are required suitably
maintenance of justified on each of the claimed specimen types.
sensitivity and/or
limit of detection,
and/or accuracy, and
precision
Specimens to Sufficient negative specimens should be included to
demonstrate ensure that the claims will be met at end of shelf life.
specificity and
validity of runs
Specimens (or If more than one part of the genome is to be detected, both
reagents) to systems must be shown to be stable.
demonstrate If both DNA and RNA are measured the complete system
stability of each of must be shown to be stable.
the critical
components of the
IVD
2
Specimens to monitor tests that measure CD4 cells
Page | 36Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
Rationale
CD4 measurements are quantitative, and accuracy at the clinical decision points is
important. The design input should have information on the accuracy and other
parameters required, and the panel must be designed to provide evidence that
these parameters are maintained over the assigned life of the reagent and
measuring IVD.
Parameters
The panel used in stability work must be able to demonstrate the following.
Stability of all the antibodies used in the IVD (frequently anti-CD4 and anti-
CD3 antibodies; any other critical components must also be covered)
Accuracy and trueness of measurement maintained at the critical level (at
least five specimens required)
Claimed linearity over the required range of CD4 count (at least five
specimens required)
Measure drift
Specimens
Artificial specimens, such as stabilized blood specimens, can be used if a risk
assessment based on R&D work indicates that they are effective. Fresh specimens
are usually required. Measurements should be compared to an approved reference
system.
Examples of approaches
Aged or in-use lots may be compared with a reference, e.g. a new lot. Precision
studies can be performed as described in CLSI guidelines.
3 Specimens to monitor tests for HIV antibodies
Specimens Remarks
Possible approaches to obtain samples :
IgM first Study the early data from commercial
seroconversion seroconversion panels where the seroconversion
specimens and IgG was frequently monitored by IgM and IgG blots
first seroconversion Study the responses to second and third generation
specimens assays or protein A and protein L assays (this
approach is less useful).
All other parts of the
HIV proteome
included, e.g. reverse
Page | 37Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
transcriptase (RT)
Late stage specimens This might serve to monitor any kit run control.
– usually a high HIV serology is not particularly genotype dependent. It is
dilution set near the usually not necessary to include controls for genotype
sample-to-cut-off ratio detection unless risk assessment or experiment shows that
it is required for a particular IVD.
HIV-2, diluted to near
the sample-to-cut-off Seroconversion specimens are very rare.
ratio
HIV-1 (0), if claimed
Difficult specimens to 100 negatives at release subject to risk analysis and
monitor specificity and statistical analysis of the allowable (relative to the claimed)
invalid rates false reactive rate and invalidity rate
4 Specimens to monitor tests for antibodies for HIV-1/2 and Treponema
pallidum (TP)
Specimens Remarks
Specimens to detect See above section 3 Specimens to monitor tests for HIV
HIV antibodies
Note: Each of these epitopes play a role in detecting
Specimens to detect syphilis in different stages of the infection. It is necessary
all the critical epitopes to have a panel member to monitor each epitope system
in the IVD, for present (and possibly each stage of infection), even if poly-
example TpN47, fusion proteins are used. This can be avoided if the
TpN17 and TpN15 manufacturer can demonstrate that each epitope system is
equally stable.
Specimens able to
show that the
invalidity and
Note: It would not be sufficient for WHO prequalification to
specificity rates do not
extrapolate to the stability of HIV-2/TP detection by testing
fall outside the claims,
only HIV-1 positive specimens.
particularly if whole
blood is a claimed
specimen type
5 Specimens to monitor tests for hepatitis C virus antibodies
Page | 38Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
Specimens Remarks
NS3 first
seroconversion
specimens and core
first seroconversion
specimens
Specimens to monitor
any other antibodies Results can be obtained from line immunoassays that
claimed (frequently differentiate antibody responses to the different proteins.
against NS5 and NS4)
Note: Hepatitis C virus serology is not particularly genotype
dependent in terms of anti-core and anti-NS3, but it is
A late stage dilution possible to make serotyping assays based on NS4 that
near the sample-to- mimic genotyping reasonably well. It is usually not
cut-off ratio necessary to include controls for genotype detection,
unless risk assessment or experiment for a particular IVD
show otherwise.
Difficult specimens to 100 negative specimens subject to risk analysis and
monitor specificity and statistical analysis of the allowable false reactive rate and
invalid rates invalidity rate (relative to the claimed rates)
6 Specimens to monitor for tests for hepatitis B surface antigen (HBsAg)
Specimens Remarks
Specimens to define Traceability is required to one of the standard of NIB or
sensitivity relative to NIBSC or WHO or any other, which are required suitably
the claim justified on each of the claimed specimen types.
Specimens to monitor
the maintenance of These will almost certainly be traceable to the “First
the claims of a variety International Reference Panel 2011, for Hepatitis B virus
of serotypes / genotype panel for HBsAg-based assays" PEI code:
genotypes and mutant 6100/09.
forms
Specimens to control
against prozone/high
Page | 39Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
dose hook effect if
found or if
theoretically an issue
If detection of HBsAg
in the presence of
anti-HBsAg is claimed
(current best practice)
proof of maintenance
of the claim
If the monoclonal antibodies used have particular function
or bias, such as against the ayr or adr serotypes (not
Specimens to monitor controlled by the standards) or to detect mutant forms of
the critical the antigen, each must be monitored to ensure viability at
components of the end of shelf life. These may be the same specimens as
IVD mentioned in the previous paragraphs.
If there are critical dissociation chemicals or red-cell
capture or rupture agents used, these must also be
monitored.
Difficult specimens to 100 negatives subject to risk analysis and statistical
monitor specificity and analysis of the allowable (relative to the claimed) false
invalid rates reactive rate and invalidity rate.
Page | 40Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
APPENDIX – III
Examples of Stability Study approaches
1 Accelerated Stability Study: (Three lots)
Product Claimed Claimed Accelerated Testing
name Storage shelf life condition interval
condition
HIV 1-2 Ab 2-30°C 24 months 2-8°C, Room 0 day, 4thday,
Rapid Temperature (20- 7thday,
qualitative test 30°C), 37°C and 14thday,
Rapid 45°C 21stday,
qualitative test 30thday and
41stday
2 Real Time Stability Study: (Three lots)
Product name Claimed Claimed Real Time Testing
Storage shelf life Stability interval
condition condition
HIV 1-2 Ab Rapid 2-30°C 24 months 2-8°C, Room 0 month (Initial
qualitative test Temperature Testing) and
Rapid qualitative (20-30°C), 3 months, up
test to 27 months in
three months
interval
Page | 41Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
3 In Use Stability Study: One lot
Product Claimed Claimed Claimed Claimed In Use Testing
name Storage shelf In Use Stability Stability or interval
condition life Stability after Open Bottle
or Open opening Stability
Bottle condition
Stability
condition
HIV 1-2 2-30°C 24 2-30°C upto 24 2-8°C, Room 0 hrs
Ab Rapid months hours Temperature (initial
qualitative after (20-30°C), testing),
test Rapid opening 37°C and then at
qualitative 45°C 0 hrs
test (initial
testing),
then at 0
hr, 1 hr,
2 hrs, 3
hrs , 4
hrs,
5hrs, 6
hrs,
24hrs
and at
25hrs
Page | 42Central Drugs Standard Control Organization
Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India
4 Shipping stability study: One lot
Product name Claimed Claimed Control Test group
Storage shelf life group temperature and
condition without Humidity cycle
temperature
and Humidity
cycle
HIV 1-2 Ab 2-30°C 24 2-30°C Cycle:1 at 40% RH
Rapid months 1-40°C for 24hrs ± 4
qualitative test Followed 30 ± 5 °C
Rapid for 24hrs ± 4
qualitative test Followed 45 ± 5 °C
for 24hrs ± 4
Followed1-40°C for
24hrs ± 4
Followed
Cycle:2 at 85% RH
1-40°C for 24hrs ± 4
Followed 30 ± 5 °C
for 24hrs ± 4
Followed 45 ± 5 °C
for 72hrs ± 4
Followed1-40°C for
24hrs ± 4
Followed
Cycle:2 at 40% RH
Followed 30 ± 5 °C
for 24hrs ± 4
Followed 45 ± 5 °C
for 24hrs ± 4
Followed1-40°C for
24hrs ± 4
Page | 43